#4: Quality Control
Contents
- Proposed by
- lemoene
- Proposal type
- Functionality
- Assigned to release
- State
- being-discussed
Definitions
Many of the concepts here are demonstrated
in the QC part of the Bika Fireassay model
Standards
A Standard is a sample of which the chemical composition is known and kept in the system
In the terminology used here, such a Standard can be used for QC purposes as
Control samples
Calibration samples
Sometimes the lab will make up it's own Standard in a container and repeatedly use some of it for analyses. Other times it will be ordered from a Standard Supplier and be delivered in a batch of say 24 ampoules, and 1 ampoule is used per analyses
The above explanation suggests that Standards AND Standard Batches in the system be treated as a Sample on which successive Analysis Requests are performed
Standards have expiry dates where after the system does not allow analyses on them
Standard Suppliers are maintained in the system per standard Bika 'organisation and contact' model with the addition of a tab listing all the Standards supplied to the lab. From this listing, links lead to the labs performance stats and graphs on that Standard
All Standards in the system can be listed off a main menu item too, leading similarly to their performance graphs
Duplicates (Checks)
Worksheets often include part of the same sample twice and the system compares their results - the exact same result must be returned for the 2 analyses in the perfect lab. Should they differ more that a specified %, the labmanager is alerted and a 'reject or retest' workflow entered
Duplicates don't have to be Standards. Most or the time labs uses any arbitrary sample on the worksheet or instrument tray
Blanks
An empty sample container is put through the analysis proses - result must be 0. Above specified % error, alarm is raised and reject or retest' workflow entered
Calibrations
Some standards are run through the analytical instrument before, and sometimes also after, the routine analyses on a tray/worksheet with the purpose of calibrating the instrument
In this design, calibration results are captured but not used as of yet
Not in the scope: In phase 2, these calibrations will be used to plot calibration graphs and if there was sufficient drift between the 1st and last analyses on the worksheet, correction factors calculated and applied
Specifications
Out of range limits are maintained to measure QC results against and labmanagers alerted when exceeded
Standards have these values stored on their individual records
Fault tolerances for Blanks and Duplicates are set per sample type and analyses in the LIMS set-up
Motivation
Required by Bika 2 sponsor
Proposal
To refine the design for Bika's QC presented here
Deliverables
Development and Implementation arising from the functional description in this document
Some of the concepts here are demonstrated
in the QC part of the Bika Fireassay model
QC Workflow
The laboratory includes a duplicates and control analyses on each worksheet. Should the results returned for it fall outside the allowable tolerance specified
the result is displayed against red background
all analyses on the same worksheet are flagged for rejection or re testing and the lab manager alerted
Reporting
QC results are looked up from a special page 'Control Analyses'
When clicking through to a control analyses, the user sees a page with the analysis' data summarised and graphed. Out of range values are displayed against shaded background
The control analysis result is also statistically compared with previous results for the Standard presented on Performance and Distribution graphs
The default period for statistical analysis is the previous 6 months' data. The user may specify any valid date range on the page and reload it - handy for looking at older analyses
These graphs and tables are also available, per sample type and analysis, from a QC reports menu
The following graphs are available
Trend - showing performance over time. true value, min & max measured and a trend plotted through the data. An example for an standard with true value of 3.00 g/t:
Distribution - shows true value, the lab's mean and normal distribution around it
An example for a result of 11.20 %/vol, lab mean of 11.15 and true value of 11.00. The standard deviation is also shown
Representations of graphs here are 'hand drawn'
To see Bika's production quality graphs
please visit the Bika Interlab page
Duplicates - Relative % error
The % error on duplicates in the results range tested. An example